According to the new, stricter provisions of the German Ordinance on Hazardous Substances, laboratories, schools and industrial facilities which work with aggressive chemical materials must be equipped with special exhaust air systems. In accordance with the DIN 1946-7/7-2009 standard the air fl ow must be monitored directly and electronically. Monitoring the ventilating fans is not enough, because blockage or malfunctions cannot be detected in this way

A company, which specialises in air technology, equipment and environmental technology as well as plastic processing, needed air fl ow sensors for their exhaust systems. Because the exhaust systems were to be used with chemically aggressive materials in particular, air fl ow sensors which are resistant to these aggressive materials were needed. They also had to be robust and easy to install in the exhaust duct.

On account of the technical specifi cations of the sensor, which also had to be resistant to sulphur-containing exhaust from galvanic baths, the company chose the Proxitron FKM 130.83 air fl ow sensor. Unlike mechanical sensors, where sensing elements (membranes or wind vanes) move and become unusable due to dirt, grease or emulsions, Proxitron air fl ow sensors are protected from wear thanks to their calorimetric functioning principle and the integrated electronics into the plastic housing.

All Proxitron air fl ow sensors feature easy installation and maintenance-free function. The 230 V AC voltage used in the system is used directly for operation and as a switching signal. An additional DC auxiliary voltage or an external evaluation device is not necessary. If the fl ow rate falls below the minimum level, the sensor interrupts the output signal. The adjustable set fl ow rate can thus give an early warning if the fl ow rate is dropping below the minimum fl ow and initiate the timely servicing of the exhaust system.

An assembly fl ange is included in the scope of supply. Alternatively suitable clamps or a M32-cable screw connection can be used for pressure-resistant assembly.  

Different cable lengths or a plug version are available

At a glance

• Corrosion-resistant plastic housing

• Adjustable fl ow limit value or easy commissioning with Teach-In function

• Maintenance-free operation

• Pressure-resistant assembly possible with M32 screw connection

• Diff erent versions available 

Technical data (FKM 130.83)

• Adjusting range: 1 - 10 m/s

• Factory setting: 3 m/s

• Ambient temperature: -10 to +60 °C

• Function display: LED

• Operating voltage: 230 V AC

• Output function: normally closed

Optionally

• Teach-In version

• Normally open version

You can find here a selection of our air flow sensors: 

24 V DC operating voltage    
Connection Output Adjustment Type
Plug S4 (M12) PNP - normally open Potentiometer or ProxiTeach FKM 230.13 G S4 / FKM 231.13 G S4
Plug S4 (M12) PNP - normally closed Potentiometer or ProxiTeach FKM 230.12 G S4 / FKM 231.12 G S4
Plug S4 (M12) NPN - normally open Potentiometer or ProxiTeach FKM 230.11 G S4
Plug S4 (M12) NPN - normally closed Potentiometer or ProxiTeach FKM 230.10 G S4
Plug S4 (M12) Analog 0-10 V   FKM 230.19 S4
Plug S4 (M12) Analog 4-20 mA   FKM 231.194 S4
Cable *) All sensors are available with a fixed connection cable in different lengths as well FKM .............. without the addition of S4

 

115 V AC operating voltage    
Connection Output Adjustment Type
Cable Normally open Potentiometer FKM 130.53
Cable Normally closed Potentiometer FKM 130.52

 

230 V AC operating voltage    
Connection Output Adjustment Type
Cable Normally open Potentiometer FKM 130.83
Cable Normally closed Potentiometer FKM 130.82